Step 3
Adjustment: log into the RMS software. Select Tools > Adjustment. Select the measuring point. Click
on <Refresh now>.
General procedure:
Type the humidity reference value and click <Acquire>
Repeat until the desired number of adjustment points are acquired.
Note: The user of this feature must take care about sufficient stable circumstances
for a data acquisition. Please take the interval into account. One value after every 60
seconds does not detect short term changes of the value.
Click <Adjust> to finish.
Use the button <A Adjust> for one point adjustment.
Click <Delete user adjustment> to delete the last customer adjustment.
Humidity 2 points acquired