Network>Network Services Settings>remove the checkmark of “Enable Device
Discovery (ADDP)”.
IMPORTANT: Parallel and unwished communication due to the opening of HW4 will be
avoided. A side effect is, that this device will not automatically be found by HW4 or the
Discovery tool anymore. To access the device via HW4, the IP address must be added
manually.
Step 6 Close connection after the following number of idle seconds:
Unwished parallel communication/requests to the Ethernet address of the Rotronic device can
cause an unwished blocking of the devices internal access to the Ethernet port. For that reason
the following setting can be done:
Serial Ports>Advanced serial settings>TCP Settings>Set a checkmark to “Close
connection after the following number of idle seconds”.
>Serial Ports> Advanced serial settings> TCP Settings> Timeout: xx seconds.
If the port is blocked then the device will renew/unblock the port by itself after the chosen timeout.
This way a permanent timeout of the device will be avoided. The loss of data will be reduced
significantly. For an RMS-CONVERTER interval of 60 seconds, a timeout of 10 to 30 seconds
would be meaningful.
Содержание RMS-CONVERTER
Страница 1: ...2017 ROTRONIC AG Bassersdorf Switzerland RMS CONVERTER...
Страница 101: ...RMS Converter E M RMS Converter V1_4 docx Instruction Manual 2018 ROTRONIC AG Bassersdorf Switzerland Page 101 of 126...
Страница 102: ...RMS Converter E M RMS Converter V1_4 docx Instruction Manual 2018 ROTRONIC AG Bassersdorf Switzerland Page 102 of 126...
Страница 106: ...RMS Converter E M RMS Converter V1_4 docx Instruction Manual 2018 ROTRONIC AG Bassersdorf Switzerland Page 106 of 126...