Step 5
Sensor error: In case of one sensor error out of the two, RMS displays a sensor error for both
measurement values.
Step 6
Measurement alarm: Alarms linked to the measurement limits and programmed within the HC2 will
not be transmitted to RMS. However if the device has a display and the display is configured
correctly, the alarm will show on the display.
Step 7
Fix values on the HC2: fixed measurement values programmed within the HC2 are displayed as
simulators within RMS.
IMPORTANT:
In case of an active fix value of one measurement value of a HC2, RMS will show
“simulator connected” for both measurement values of the probe.