Rosemount Model P330 Pressure Validator
4-8
Setting units in any task mode other than Basic or Advanced
1. Press the
OUTPUT
key.
2. Use
¶
and
Ä
to highlight
UNITS
. Press
ENTER
.
3. Use
¶
and
Ä
to highlight desired units. Press
ENTER
.
4. If desired units are not listed, refer to Selecting Units on page
4-4.
Setting units in Basic mode
1. Press
F2
.
2. Use
¶
and
Ä
to highlight
UNITS
. Press
ENTER
.
3. Use
¶
and
Ä
to highlight desired units. Press
ENTER
.
4. If desired units are not listed, refer to Selecting Units on page
4-4.
Setting units in Advanced mode
1. Press the
OUTPUT
key.
2. Press
F2
.
3. Use
¶
and
Ä
to highlight
UNITS
. Press
ENTER
.
4. Use
¶
and
Ä
to highlight desired units. Press
ENTER
.
5. If desired units are not listed, see Selecting Units on page 4-4.
Zeroing the Display
Both the input and the output displays should be set to zero before a
validation.
NOTE
If the display is more than 5% of zero and an attempt is made to zero it,
the screen displays the message
ZERO TOO LARGE
, and the display will
not zero. Turn the release valve counterclockwise to make sure the
system is depressurized.
To zero the input and output displays, press the shaded keys in Figure
4-10, and proceed as follows:
FIGURE 4-10. Zeroing the Input and
Output Displays.
P
3
30
_5
4
A
Содержание P330
Страница 1: ...Model P330 Pressure Validator Installation and Operation Manual 00809 0100 4776 English Rev AA ...
Страница 2: ......
Страница 4: ...Rosemount Model P330 Pressure Validator ...
Страница 8: ...iv ...
Страница 12: ...Rosemount Model P330 Pressure Validator 1 4 ...
Страница 20: ...Rosemount Model P330 Pressure Validator 2 8 ...
Страница 30: ...Rosemount Model P330 Pressure Validator 3 10 ...
Страница 40: ...Rosemount Model P330 Pressure Validator 4 10 ...
Страница 68: ...Rosemount Model P330 Pressure Validator 6 10 ...
Страница 80: ...Rosemount Model P330 Pressure Validator 8 6 ...
Страница 91: ... 3 ...