Appendix A
Reference Data
Functional Specifications . . . . . . . . . . . . . . . . . . . . . . . . . page A-1
Foundation™ fieldbus Specifications . . . . . . . . . . . . . . . page A-4
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . page A-5
Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-7
FUNCTIONAL
SPECIFICATIONS
Sensor Compatibility
Compatible with Rosemount 8705, 8711, 8721, and 570TM sensors.
Compatible with Rosemount 8707 sensor with D2 Dual calibration option.
Compatible with AC and DC powered sensors of other manufacturers.
Sensor Coil Resistance
350
maximum
Flow Rate Range
Capable of processing signals from fluids that are traveling between 0.04 and
39 ft/s (0.01 to 12 m/s) for both forward and reverse flow in all sensor sizes.
Full scale continuously adjustable between –39 and 39 ft/s (–12 to 12 m/s).
Conductivity Limits
Process liquid must have a conductivity of 5 microsiemens/cm (5
micromhos/cm) or greater for 8732E. Excludes the effect of interconnecting
cable length in remote mount transmitter installations.
Power Supply
90 -250 V AC ±10%, 50–60 Hz or 12-42 V DC
AC Power Supply Requirements
Units powered by 90-250 V AC have the following power requirements.
Figure A-1. AC Current
Requirements
Supply C
urr
ent (
A
mps)
0.100
0.120
0.140
0.180
0.200
0.220
0.240
0.260
0.280
0.300
0.320
0.160
80
100
120
140
160
180
200
220
240
Power Supply Voltage (AC RMS)
Содержание 8732 Series
Страница 2: ......
Страница 4: ......
Страница 10: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 TOC 6 ...
Страница 26: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 2 14 ...
Страница 38: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 3 12 ...
Страница 92: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 A 8 ...
Страница 107: ...Reference Manual 00809 0100 4663 Rev BA January 2010 B 15 Rosemount 8732 Figure B 1 ATEX Installation 1 of 6 ...
Страница 108: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 B 16 Figure B 2 ATEX Installation 2 of 6 ...
Страница 109: ...Reference Manual 00809 0100 4663 Rev BA January 2010 B 17 Rosemount 8732 Figure B 3 ATEX Installation 3 of 6 ...
Страница 110: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 B 18 Figure B 4 ATEX Installation 4 of 6 ...
Страница 111: ...Reference Manual 00809 0100 4663 Rev BA January 2010 B 19 Rosemount 8732 Figure B 5 ATEX Installation 5 of 6 ...
Страница 112: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 B 20 Figure B 6 ATEX Installation 6 of 6 ...
Страница 113: ...Reference Manual 00809 0100 4663 Rev BA January 2010 B 21 Rosemount 8732 Figure B 7 FM Certified I S Output 1 of 4 ...
Страница 114: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 B 22 Figure B 8 FM Certified I S Output 2 of 4 ...
Страница 115: ...Reference Manual 00809 0100 4663 Rev BA January 2010 B 23 Rosemount 8732 Figure B 9 FM Certified I S Output 3 of 4 ...
Страница 116: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 B 24 Figure B 10 FM Certified I S Output 4 of 4 ...
Страница 117: ...Reference Manual 00809 0100 4663 Rev BA January 2010 B 25 Rosemount 8732 Figure B 11 CSA Certified I S Output 1 of 2 ...
Страница 118: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 B 26 Figure B 12 CSA Certified I S Output 2 of 2 ...
Страница 119: ...Reference Manual 00809 0100 4663 Rev BA January 2010 B 27 Rosemount 8732 Figure B 13 CSA Installation ...
Страница 120: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 B 28 Figure B 14 Factory Mutual Hazardous Locations ...
Страница 186: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 H 8 ...
Страница 189: ......