![Rosemount 8732 Series Скачать руководство пользователя страница 135](http://html1.mh-extra.com/html/rosemount/8732-series/8732-series_reference-manual_1487411135.webp)
Reference Manual
00809-0100-4663, Rev BA
January 2010
C-15
Rosemount 8732
Electrode Circuit Resistance
The Electrode Circuit Resistance is a measurement of the electrode circuit
health. This value is used as a baseline to determine if the electrode circuit is
still operating correctly when the 8714i Meter Verification diagnostic is
initiated.
8714i Meter Verification Measurements
The 8714i Meter Verification test will make measurements of the coil
resistance, coil signature, and electrode resistance and compare these values
to the values taken during the sensor signature process to determine the
sensor calibration deviation, the coil circuit health, and the electrode circuit
health. In addition, the measurements taken by this test can provide additional
information when troubleshooting the meter.
Coil Circuit Resistance
The Coil Circuit Resistance is a measurement of the coil circuit health. This
value is compared to the coil circuit resistance baseline measurement taken
during the sensor signature process to determine coil circuit health.
Coil Signature
The Coil Signature is a measurement of the magnetic field strength. This
value is compared to the coil signature baseline measurement taken during
the sensor signature process to determine sensor calibration deviation.
Electrode Circuit Resistance
The Electrode Circuit Resistance is a measurement of the electrode circuit
health. This value is compared to the electrode circuit resistance baseline
measurement taken during the sensor signature process to determine
electrode circuit health.
375
Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Sensor Signature, Signature Values, Electrode Resistance
AMS Tab
Config/Setup, 8714i
375
Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Measurements, Coil Resistance
AMS Tab
Config/Setup, 8714i
375
Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Measurements, Coil Signature
AMS Tab
Config/Setup, 8714i
375
Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Measurements, Electrode Resistance
AMS Tab
Config/Setup, 8714i
Содержание 8732 Series
Страница 2: ......
Страница 4: ......
Страница 10: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 TOC 6 ...
Страница 26: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 2 14 ...
Страница 38: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 3 12 ...
Страница 92: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 A 8 ...
Страница 107: ...Reference Manual 00809 0100 4663 Rev BA January 2010 B 15 Rosemount 8732 Figure B 1 ATEX Installation 1 of 6 ...
Страница 108: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 B 16 Figure B 2 ATEX Installation 2 of 6 ...
Страница 109: ...Reference Manual 00809 0100 4663 Rev BA January 2010 B 17 Rosemount 8732 Figure B 3 ATEX Installation 3 of 6 ...
Страница 110: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 B 18 Figure B 4 ATEX Installation 4 of 6 ...
Страница 111: ...Reference Manual 00809 0100 4663 Rev BA January 2010 B 19 Rosemount 8732 Figure B 5 ATEX Installation 5 of 6 ...
Страница 112: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 B 20 Figure B 6 ATEX Installation 6 of 6 ...
Страница 113: ...Reference Manual 00809 0100 4663 Rev BA January 2010 B 21 Rosemount 8732 Figure B 7 FM Certified I S Output 1 of 4 ...
Страница 114: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 B 22 Figure B 8 FM Certified I S Output 2 of 4 ...
Страница 115: ...Reference Manual 00809 0100 4663 Rev BA January 2010 B 23 Rosemount 8732 Figure B 9 FM Certified I S Output 3 of 4 ...
Страница 116: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 B 24 Figure B 10 FM Certified I S Output 4 of 4 ...
Страница 117: ...Reference Manual 00809 0100 4663 Rev BA January 2010 B 25 Rosemount 8732 Figure B 11 CSA Certified I S Output 1 of 2 ...
Страница 118: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 B 26 Figure B 12 CSA Certified I S Output 2 of 2 ...
Страница 119: ...Reference Manual 00809 0100 4663 Rev BA January 2010 B 27 Rosemount 8732 Figure B 13 CSA Installation ...
Страница 120: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 B 28 Figure B 14 Factory Mutual Hazardous Locations ...
Страница 186: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 H 8 ...
Страница 189: ......