S
ERVICE AND
M
AINTENANCE
748213-P Rosemount Analytical June 1997
7-7
Model 755R Oxygen Analyzer
Red Mark for
Alignment
1/4"
Set Screw
C8
C1
R10
Voltmeter
Lead
R8
R7
DETAIL A
DETAIL B
Place a digital voltmeter between the wiper of zero potentiometer (R13) and TP7 (ground).
Adjust for 0 VDC.
Remove the voltmeter from R13 and place on R10 (see Figure 7-4B). Adjust R9 for 0 VDC.
Remove the voltmeter from R10 and place on TP8. Move the photocell to obtain a DC
voltage as close to 0 mV as possible, but no more than
±
750 mV.
Apply power to instrument and allow to warm-up approximately one hour.
Set front panel ZERO at mid-range (i.e., five turns from either end).
Connect digital voltmeter from slider of R9 to chassis ground. With a steady flow of 50 to
500 cc/min of nitrogen zero gas going through instrument, adjust R9 for 0 V.
Connect the voltmeter between TP10 and circuit ground (TP7). Adjust front panel ZERO
for reading of exactly zero on voltmeter.
All internal adjustments are now properly set. The instrument may be calibrated per
Section 3.4.
F
IGURE
7-4. L
AMP
R
EPLACEMENT