S
ERVICE AND
M
AINTENANCE
7-10
June 1997 Rosemount Analytical 748213-P
Model 755R Oxygen Analyzer
R
ECORDER
F
ULLSCALE
1. Flow nitrogen at 250 cc/min, monitor TP16, and adjust front panel ZERO
potentiometer for .000 VDC.
2. Flow 100% oxygen for span gas. Recorder output for 1 V, 100 mV, or 10 mV
should read 100% of span gas. Adjust R88 if necessary.