T
HEORY
748213-P Rosemount Analytical June 1997
5-7
Model 755R Oxygen Analyzer
Potentiometric output is strap-selectable for 0 to 10 mV, 0 to 100 mV, 0 to 1 V, or 0 to
5VDC. Potentiometer R88 permits adjustment of recorder span on 0 to 1 V, 100 mV
and 10 mV outputs.
5.3.3 P
OWER
S
UPPLY
B
OARD
A
SSEMBLY
The Power Supply Board assembly contains power supply and temperature control
circuitry. The assembly is mounted within the analyzer case.
As shown in DWG 617186, the various circuits operate on main power transformer T1.
During instrument assembly, the two primary windings of T1 are factory-connected for
operation on either 115 VAC or 230 VAC, as noted on the name rating plate.
The same circuit board contains the following:
S
OURCE
L
AMP
P
OWER
S
UPPLY
S
ECTION
This circuit provides a regulated output of 2.20 VDC to operate incandescent source
lamp DS1 within the optical bench assembly. One secondary of main power
transformer T1 drives a fullwave rectifier consisting of CR7 and CR8. The output of
DS1 is held constant by a voltage regulator circuit utilizing U7, Q4 and Q5.
T
HE
±±±±
15 V P
OWER
S
UPPLY
S
ECTION
This section provides DC voltage required for various amplifiers and other circuits.
Fullwave rectifier bridge CR5 provides both positive and negative outputs. Each is
routed through an associated series type integrated circuit, voltage regulator, providing
regulated outputs of +15 V and -15 V.
D
ETECTOR
T
EMPERATURE
C
ONTROL
S
ECTION
This section maintains the detector at a controlled temperature of 150
°
F (66
°
C).
Temperature is sensed by RT1, a resistance element permanently attached to the
detector assembly. The signal from the sensor is applied to amplifier AR6, which
drives transistors Q2 and Q3, thus controlling application of DC power from full wave
rectifier bridge CR6 to two heaters within the detector/magnet assembly: HR1,
mounted on the top of the magnet and HR2, permanently mounted on the rear of the
detector assembly.
D
ETECTOR
C
OMPARTMENT
T
EMPERATURE
C
ONTROL
S
ECTION
This section maintains the interior of the detector compartment at a controlled
temperature of 140
°
F (60
°
C). Temperature is sensed by a thermistor located in the
detector compartment and plugged into the Control Board assembly.
The circuit provides an on-off control of heater element HR3 via TRIAC element Q7.
Heater HR3 is a part of the heater/fan assembly.
Содержание 755R
Страница 1: ...Rosemount Analytical MODEL 755R OXYGEN ANALYZER INSTRUCTION MANUAL ...
Страница 8: ...CONTENTS vi June 1997 Rosemount Analytical 748213 P Model 755R Oxygen Analyzer NOTES ...
Страница 62: ...CIRCUIT ANALYSIS 6 14 June 1997 Rosemount Analytical 748213 P Model 755R Oxygen Analyzer NOTES ...
Страница 76: ... blank ...
Страница 78: ... blank ...
Страница 80: ... blank ...