S
ERVICE AND
M
AINTENANCE
748213-P Rosemount Analytical June 1997
7-5
Model 755R Oxygen Analyzer
Side View
of Connector
Keeper
Connector Pin Removed
Upper Slot
Lower Slot
Connector Pin/
Leads in Place
Improvised Pin Removal Tool, Such as a Paper Clip
When inserting a pin/lead, its keeper must face toward the slot opening in the
connector in order to lock in. If inserted otherwise, the pin/lead will be forced out when
the two connectors are joined.
F
IGURE
7-2. P
IN
/L
EAD
R
EMOVAL
F
IGURE
7-3. D
ETECTOR
O
PTICAL
B
ENCH
7.3.1 S
OURCE
L
AMP
The simplest check of the source lamp is to verify that it is lit. Another check is done
by removing the housing cover and viewing the lamp through the photocell alignment
hole (see Figure 7-3). If the photocell is not illuminated, test the voltage across TP2 to
TP5 (ground). This voltage should be 2.2 V
±
0.2 VDC. If reading is correct, the lamp
may be burned out; also inspect the cable for continuity. If voltage reading is not 2.2
V
±
0.2 VDC, the Power Supply Board must be replaced.
Hole for Source Lamp
When dual photocell is installed,
the gap between the two
photocells should be in position
indicated by this line.
Optical Bench
10
J12
1
HR2
Suspension
Heater
PUR
GRN
18
BLK
BLK
WHT
WHT
RT1
Suspension
Terminals
Содержание 755R
Страница 1: ...Rosemount Analytical MODEL 755R OXYGEN ANALYZER INSTRUCTION MANUAL ...
Страница 8: ...CONTENTS vi June 1997 Rosemount Analytical 748213 P Model 755R Oxygen Analyzer NOTES ...
Страница 62: ...CIRCUIT ANALYSIS 6 14 June 1997 Rosemount Analytical 748213 P Model 755R Oxygen Analyzer NOTES ...
Страница 76: ... blank ...
Страница 78: ... blank ...
Страница 80: ... blank ...