I
NSTALLATION
748213-P Rosemount Analytical June 1997
2-9
Model 755R Oxygen Analyzer
I/O board.
3. Connect the output cable to the appropriate terminals of the recorder or other
potentiometric device:
a. For device with span of 0 to 10 mV, 0 to 100 mV, 0 to 1V, or 0 to 5V, connect
cable directly to input terminals of the device, ensuring correct polarity and
range selection.
b. For a device with intermediate span (i.e., between the specified values),
connect the cable to the device via a suitable external voltage divider (See
Figure 2-3).
2.4.2.2 I
SOLATED
C
URRENT
O
UTPUT
(O
PTIONAL
)
1. Verify that the optional current output board appropriate to desired output is
properly in place in its connector. See Figure 3-1. If originally ordered with the
analyzer, the board is factory installed.
2. On I/O board, connect leads of shielded recorder cable to “CURRENT OUT+” and
“-” terminals.
3. Connect free end of output cable to input terminals of recorder or other
current-actuated device, making sure that polarity is correct. If two or more
current-actuated devices are to be used, they must be connected in series (See
Figure 2-4). Do not exceed the maximum load resistance of 1000 ohms.
Current and voltage outputs may be utilized simultaneously if desired.
F
IGURE
2-4. M
ODEL
755R C
ONNECTED TO
D
RIVE
S
EVERAL
C
URRENT
-A
CTUATED
O
UTPUT
D
EVICES
Note: Total series resistance of all devices is not to exceed 1000 ohms.
755R
Analyzer
Recorder
Controller
Remote
Indicator
+
-
+
-
+
-
+
-
mA
Содержание 755R
Страница 1: ...Rosemount Analytical MODEL 755R OXYGEN ANALYZER INSTRUCTION MANUAL ...
Страница 8: ...CONTENTS vi June 1997 Rosemount Analytical 748213 P Model 755R Oxygen Analyzer NOTES ...
Страница 62: ...CIRCUIT ANALYSIS 6 14 June 1997 Rosemount Analytical 748213 P Model 755R Oxygen Analyzer NOTES ...
Страница 76: ... blank ...
Страница 78: ... blank ...
Страница 80: ... blank ...