748213-P Rosemount Analytical June 1997
7-9
Model 755R Oxygen Analyzer
7.5 CONTROL BOARD SETUP
P
OWER
S
UPPLY
T
EST
1. TP7 (circuit ground) is ground point for all voltage tests.
2. Counterclockwise end of front panel ZERO potentiometer (R13 on Control Board):
-15 VDC
±
0.5 VDC.
3. Clockwise end of ZERO potentiometer: +15 VDC
±
5 VDC.
4. Set ZERO potentiometer to obtain a reading of .0 VDC
±
10 mV at slider.
5. Measure TP19: +5 VDC
±
0.25 VDC.
D
ETECTOR ZERO
1. Flow 250 cc/min nitrogen.
2. Monitor TP8, adjust R9 for 0 VDC
±
2mV.
U4 Z
ERO
1. Monitor TP5, adjust R100 for 1 VDC
±
2mV.
2. Monitor TP10, adjust R13 (ZERO) for 0.0 VDC
±
5mV.
U8 Z
ERO
Monitor TP11, adjust R29 for 0.0 VDC
±
5mV.
U10 Z
ERO
1. Monitor TP16, adjust R29 for 0.0 VDC
±
5mV.
Note:
This adjustment requires a "long time" constant. Allow adequate time.
2. Adjust R29 to obtain a reading of 00.00
±
5 counts on the display.
F
ULLSCALE
1. Monitor TP11. Flow 100% oxygen, adjust SPAN potentiometer R20 for 10.00 V
±
5mV.
2. Monitor TP16, adjust R20 for 10.00 V
±
5mV. The display must read 100.00
±
5
counts.