If resistance is correct, and the combined resistance is incorrect, heater HR1 may be
open.
To reach the leads of HR1, remove the circuit board on the heater assembly.
Resistance of HR1 should be approximately 21 ohms.
To check operation of the heater circuit, connect a voltmeter across R61 on the Power
Supply Board. Normally, the voltage will be 4 VDC when cold and will drop to
approximately 0.4 VDC at control temperature. Temperature sensor RT1 is mounted
in the detector, with leads accessible at pins 10 and 11 of detector connector J12.
The sensor resistance should be 1M ohms at 25
°
C and approximately 149K ohms at
operating temperature of 65
°
C.
F
IGURE
7-1.
D
ETECTOR
/M
AGNET
A
SSEMBLY
7.3 DETECTOR CHECK
To isolate the detector as the problem, it is necessary to check the source lamp,
photocells, and suspension (see Figure 7-1B). These components are connected via
J12 on the optical bench assembly.
Pin/leads may be removed from connector J12 by use of an improvised pin removal
tool, such as a paper clip (see Figure 7-2).
Connect J12 has slots at top and bottom. To remove a connector pin/lead, insert the
tool into the upper or lower slot and push down on the end to release the keeper on
the pin.
Dual
Photocell
Connector
J12
Lamp Retaining
Set Screw
Photocell
Lock Screws (2)
Lamp Viewing
Hole
Source Lamp
Assembly
Mounting Screws (2)
Optical Bench Assembly
Detector Assembly
Magnet
Assembly
Sample
Outlet Tube
Sample Inlet
Tube
Sample Pre-Heating
Coil
Connector
Board
A. Detector/Magnet Assembly - Exploded View B. Optical Bench - Exploded View
Содержание 755R
Страница 1: ...Rosemount Analytical MODEL 755R OXYGEN ANALYZER INSTRUCTION MANUAL ...
Страница 8: ...CONTENTS vi June 1997 Rosemount Analytical 748213 P Model 755R Oxygen Analyzer NOTES ...
Страница 62: ...CIRCUIT ANALYSIS 6 14 June 1997 Rosemount Analytical 748213 P Model 755R Oxygen Analyzer NOTES ...
Страница 76: ... blank ...
Страница 78: ... blank ...
Страница 80: ... blank ...