245364-U Rosemount Analytical April 2000
6-13
Model 755A Oxygen Analyzer
A separate regulator circuit, which operates from the +15VDC supply, provides a
regulated 5VDC for the digital functions associated with the display.
6.9 ANALOG OUTPUT CIRCUITS FOR RECORDER AND
ALARMS
Refer to Figure 6-10. The analog output circuits utilize two amplifiers:
F
IRST
S
TAGE
A
MPLIFIER
This amplifier permits selecting the desired fullscale oxygen range for the recorder by
an appropriate combination of scale expansion and zero suppression.
Scale expansion is accomplished by selecting the appropriate feedback resistor in a
switch-selectable network, thus establishing one of six values for amplifier gain:
DESIRED
OXYGEN SPAN
FOR RECORDER
REQUIRED
AMPLIFIER
GAIN
100%
1X
20%
5X
10%
10X
5%
20X
2%
50X
1%
100X
The desired zero suppression is obtained as the sum of (a) a jumper-selectable fixed
value of 0%, 20%, 40%, or 80% oxygen and (b) a continuously adjustable value of 0%
to 25% oxygen. Thus the total zero suppression may be set for any desired valued
from 0% to 105% oxygen.
In order to establish the precise zero suppression required for the recorder output, the
actual applied zero suppression may be read on the digital display by placing the front
panel TEST Switch in position 4 and the recorder scale-expansion jumper in the 100%
(i.e., 1X gain) position.
Note that, in this mode the signal input is disconnected, allowing only the input offset
current for the zero-suppression function to flow into the amplifier summing node.
Example 1, Selection of a Zero Based Recorder Oxygen Range:
Desired oxygen range for recorder output: 0% to 100%
Required span is 100% oxygen; thus a gain of 1 is selected with the jumper.
Required zero suppression is 0%; thus, the Zero Suppression Selection Jumper
is removed, and the Zero Suppression Potentiometer is set for 0%.
Содержание 755A
Страница 1: ...Rosemount Analytical MODEL 755A OXYGEN ANALYZER INSTRUCTION MANUAL 245364 U ...
Страница 16: ...PREFACE P 8 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 34: ...UNPACKING AND INSTALLATION 2 14 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 50: ...ROUTINE OPERATION 4 2 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 76: ...ELECTRONIC CIRCUIT ANALYSIS 6 16 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 86: ...ROUTINE SERVICE AND MAINTENANCE 7 10 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 90: ... blank ...
Страница 92: ... blank ...
Страница 94: ... blank ...