E
LECTRONIC
C
IRCUIT
A
NALYSIS
245364-U Rosemount Analytical April 2000
6-15
Model 755A Oxygen Analyzer
F
IGURE
6-10. S
IMPLIFIED
A
NALOG
O
UTPUT
C
IRCUIT FOR
R
ECORDER
(S
HOWING
T
HREE
R
ANGES
)
10
20K
.1uF
1
To Recorder
-
+
U13
Front
Panel
TEST
100X
50X
20X
1X
-
+
U16
5V
.1uF
10X
5X
2M
To Alarm and V/I
Amplifier Gain
(Jumper Selectable)
Recorder Output
(Jumper Selectable)
20K
R50A
20K
1M
200K
100K
20K
2
3
4
Recorder
Zero
Suppression
(Jumper
Selectable)
Continuously
Variable
Recorder
Zero
Suppression
(0% to 25%
Oxygen)
+REF
-REF
20K
75K
17.8K
25K
80%
8.33K
60%
25K
25K
25K
40%
20%
1V
100mV
10mV
Resistor Values are in Ohms.
Содержание 755A
Страница 1: ...Rosemount Analytical MODEL 755A OXYGEN ANALYZER INSTRUCTION MANUAL 245364 U ...
Страница 16: ...PREFACE P 8 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 34: ...UNPACKING AND INSTALLATION 2 14 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 50: ...ROUTINE OPERATION 4 2 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 76: ...ELECTRONIC CIRCUIT ANALYSIS 6 16 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 86: ...ROUTINE SERVICE AND MAINTENANCE 7 10 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 90: ... blank ...
Страница 92: ... blank ...
Страница 94: ... blank ...