3.5 DUAL ALARM OPTION
3.5.1 I
NITIAL
C
ALIBRATION AND
S
ELECTION OF
S
ETPOINTS FOR
A
LARMS
The ALARM 1 and ALARM 2 circuits have independent setpoint and deadband
adjustments (Table 3-1, Figure 3-2). Initially the ALARM 1 and ALARM 2 setpoint
adjustments must be calibrated by means of the ALARM 1 and ALARM 2 calibration
adjustments.
1. Set TEST switch to NORM.
2. Pass upscale standard gas through the analyzer at a flow rate of 50 to 500 cc/min.
3. Verify that ALARM 1 and ALARM 2 deadband adjustments are fully
counterclockwise to set deadband at minimum. Normally these potentiometers are
factory-set for minimum deadband. Both potentiometers must remain at this
setting throughout calibration of the alarm setpoints.
C
ALIBRATION OF
ALARM 1, HIGH:
1. Rotate setpoint adjustment R64 full counterclockwise.
2. Adjust SPAN control to obtain a display or recorder reading exactly fullscale. If the
fullscale setting cannot be reached, then set to a reading higher than the desired
alarm setpoint.
3. Set ALARM 1 calibration adjustment, R63, to its clockwise limit. Turn R63
counterclockwise the minimum required to energize ALARM 1 (relay K1, Figure 3-
4). Energization may be verified by connecting an ohmmeter to relay terminals on
638254 Alarm Relay Assembly (Figure 2-1).
C
ALIBRATION OF
ALARM 2,
LOW:
1. Rotate setpoint potentiometer R67 full counterclockwise.
2. Adjust SPAN control to obtain a display or recorder reading exactly fullscale. . If the
fullscale setting cannot be reached, then set to a reading higher than the desired alarm
setpoint.
3. Set ALARM 1 calibration adjustment, R67, to its clockwise limit. Turn R63
counterclockwise the minimum required to energize ALARM 1 (relay K1, Figure 3-4).
Energization may be verified by connecting an ohmmeter to relay terminals on 638254
Alarm Relay Assembly (Figure 2-1).
S
ETPOINT OF
ALARM 1, HIGH
1. With span gas flowing, adjust SPAN Control to read desired alarm setpoint on
display or recorder.
2. Turn setpoint adjustment, R64, clockwise to energy relay.
Содержание 755A
Страница 1: ...Rosemount Analytical MODEL 755A OXYGEN ANALYZER INSTRUCTION MANUAL 245364 U ...
Страница 16: ...PREFACE P 8 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 34: ...UNPACKING AND INSTALLATION 2 14 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 50: ...ROUTINE OPERATION 4 2 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 76: ...ELECTRONIC CIRCUIT ANALYSIS 6 16 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 86: ...ROUTINE SERVICE AND MAINTENANCE 7 10 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 90: ... blank ...
Страница 92: ... blank ...
Страница 94: ... blank ...