April 2000 Rosemount Analytical 245364-U
Model 755A Oxygen Analyzer
R9 for 0 VDC.
c. Place the voltmeter on TP8 and TP7, move the photocell to obtain a DC voltage
as close to 0mV as possible but not more than
±
750mV.
4. Perform steps 8 through 11 in Section 7.3.1.
With all internal adjustments now properly set, the instrument may be calibrated in the
normal manner.
7.4 HEATING CIRCUITS
To ensure against damage from overheating in the event of malfunction, the heating
circuits are protected by thermal fuses F2 and F3. If temperature of a heated area
exceeds the permissible maximum, the associated fuse melts, opening the circuit.
Note
Thermal fuses F2 and F3 are to be plugged in, never soldered. If soldered, the
fuse element may melt and open the circuit.
7.4.1 C
ASE
H
EATER
C
ONTROL
C
IRCUIT
The case heater control circuit receives power via thermal fuse F2 (76
°
C). This fuse,
located on the Case Board, can be checked for continuity.
Case heater element HR3, located on the heater/fan assembly, has a normal
resistance of 20 ohms.
To verify heater operation, place a hand beside the right-hand side of the detector
housing. Heated air should be felt. If it is not, check the case heating circuit.
Temperature sensor RT1 has a cold resistance of 22.7K ohms and a normal operating
resistance of 20.2K ohms, indicating normal operating temperature of 140
°
F (60
°
C).
Until thyristor RV1 reaches operating temperature, it bypasses most of the current that
would otherwise flow through Triac Q7.
As a further check, disconnect connector plug P8 on the Control Board, thus
disconnecting temperature sensor RT1. Substitute a decade resistor box to simulate
the resistance of RT1. Also, connect an AC voltmeter from the hot side of the line to
the neutral side of F2.
Set the decade box for 20.2K ohms to simulate RT1 resistance at controlling
temperature. The voltmeter should now show pulses of 1 VAC.
Содержание 755A
Страница 1: ...Rosemount Analytical MODEL 755A OXYGEN ANALYZER INSTRUCTION MANUAL 245364 U ...
Страница 16: ...PREFACE P 8 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 34: ...UNPACKING AND INSTALLATION 2 14 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 50: ...ROUTINE OPERATION 4 2 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 76: ...ELECTRONIC CIRCUIT ANALYSIS 6 16 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 86: ...ROUTINE SERVICE AND MAINTENANCE 7 10 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 90: ... blank ...
Страница 92: ... blank ...
Страница 94: ... blank ...