245364-U Rosemount Analytical April 2000
6-3
Model 755A Oxygen Analyzer
-15V
R74
590K
-
+3
R75
210K
R76
37.4K
C40
2200uF
C37
1.0uF
Q6
R80
10K
R79
10K
R81
56.2
C38
.18uF
R87
10K
R78
249K
R77
10K
TO
COMPARATOR
R83
63.4K
R84
169K
R82
9.09K
RT1
-15V
-15V to 1.88V
±
0.3V
+15V
+15V
-15V
INPUT FROM
MULTIVIBRATOR
-15V
OFF
OFF
OFF
-2.3V
+2.3V
6 Hz
T2
An approximate 8V peak-to-peak AC signal is applied to comparators 1 and 2. As the
signal starts going positive, comparator 2 transistor ceases conducting and
comparator 1 transistor is off. When the signal exceeds the +159 mV on the
non-inverting terminal, it turns on comparator 1 and the output is -15V.
Comparator 1 stays on until the signal drops below +159 mV, at which time the output
will be the value of the OR bus. As the AC signal goes negative with respect to
ground, the transistor of comparator 2 conducts and the output is again -15V. The
output remains at -15VDC until the incoming signal crosses zero value and the
positive signal causes the comparator 2 transistor to cease to conduct.
Summing the effects of the two comparators in the OR circuit results in no output from
the comparators for about 4
°
of the sinewave, 2
°
after the signal goes positive (0 to 2
°
)
and 2
°
before the positive signal reaches 180
°
(178
°
to 180
°
).
During the period that neither comparator is conducting, the value on the OR bus is the
potential from the temperature-sensing bridge plus the effect of the ramp generator,
probably -1.88
±
0.03V.
The on-off effect of the comparators to the OR circuit results in application of a
positive-going pulse (from -15V to -1.89V) to the temperature bridge at the rate of 120
pulses per second.
Capacitor C36 is added to the input circuit to delay the incoming AC signal so that the
pulses will occur at or just after the line frequency crossover point.
Circuits for a ramp generator (Figure 6-2) and a temperature-sensing bridge are part of the
case heater control circuit (Figures 6-3 and 6-4).
F
IGURE
6-2. R
AMP
G
ENERATOR
Содержание 755A
Страница 1: ...Rosemount Analytical MODEL 755A OXYGEN ANALYZER INSTRUCTION MANUAL 245364 U ...
Страница 16: ...PREFACE P 8 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 34: ...UNPACKING AND INSTALLATION 2 14 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 50: ...ROUTINE OPERATION 4 2 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 76: ...ELECTRONIC CIRCUIT ANALYSIS 6 16 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 86: ...ROUTINE SERVICE AND MAINTENANCE 7 10 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 90: ... blank ...
Страница 92: ... blank ...
Страница 94: ... blank ...