b. Pass upscale standard gas through analyzer at same flow rate as was used for
downscale standard gas. Allow gas to purge analyzer for minimum of 3
minutes.
c. Adjust SPAN Control so that reading on digital display or recorder is
appropriate to the upscale standard gas. (The required reading may be the
actual oxygen content of the upscale standard gas or may be an adjusted
value, depending on the relative magnetic susceptibilities involved. Refer to
Section 3.4.2). If proper reading is unobtainable by adjustment of the SPAN
control, refer to Section 6, Routine Servicing.
Repeat steps 1 and 2 to ensure no interaction has occurred.
3.3.4 C
ALIBRATION OF
A
UTOMATIC
P
RESSURE
C
OMPENSATION
Oxygen readout is automatically corrected to within
±
1% of fullscale for barometric
pressure variations within
±
3% of the target value and is corrected to within
±
2% of
fullscale for variations within
±
5% of the target value. The target value may be set
anywhere within the range of -2.7 to 3.3 psig
±
3 psig (-18.6 to 22.8 kPa
±
21 kPa). The
factory setting is 0 psig (0 kPa). This setting is suitable for applications where (a) the
analyzer exhaust port is vented directly to the atmosphere, and (b) the installation site
is at or near sea level. If conditions are otherwise, the pressure compensation circuit
must be re-calibrated by the following procedure:
F1
Compound Range
Pressure Gauge or
Mercury Manometer
Pressure
Gauge
Vacuum
Pump
PR1
Sample
In
Sample In
755A
250
cc/min
V1
V3
10 psig
V2
V4
F
IGURE
3-3. C
ALIBRATION BY
P
RESSURE
D
ECREASE
S
ETUP
Содержание 755A
Страница 1: ...Rosemount Analytical MODEL 755A OXYGEN ANALYZER INSTRUCTION MANUAL 245364 U ...
Страница 16: ...PREFACE P 8 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 34: ...UNPACKING AND INSTALLATION 2 14 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 50: ...ROUTINE OPERATION 4 2 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 76: ...ELECTRONIC CIRCUIT ANALYSIS 6 16 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 86: ...ROUTINE SERVICE AND MAINTENANCE 7 10 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 90: ... blank ...
Страница 92: ... blank ...
Страница 94: ... blank ...