U
NPACKING AND
I
NSTALLATION
245364-U Rosemount Analytical April 2000
2-13
Model 755A Oxygen Analyzer
Recommended supply pressure is 20 psig (138 kPa), which provides a flow of
approximately 8 cubic feet per hour (4 liters per minute) and a case pressure of
approximately 0.2 inch of water (50 Pa).
With a flow rate of 4 liters per minute, four case volumes of purge gas pass through
the case in 10 minutes.
All conduit connections through the case must be sealed thoroughly with the sealant
supplied in the kit. The sealant (applied from the interior of the case) must thoroughly
cover all existing leads, as well as the conduit fitting.
F
IGURE
2-8. I
NSTALLATION OF
P
URGE
K
IT
(O
PTIONAL
)
645835 Purge
Outlet Fitting
190697 Purge
Inlet Fitting
Flow
Indicator
Purge Supply
Analyzer
Door
Affix Warning
Label
190697 Purge
Inlet Fitting
645835 Purge
Outlet Fitting
Pressure
Indicator or
Alarm
Purge
Supply
Analyzer
Door
Affix Warning
Label
A. Option with Flow Indicator B. Option with Pressure Indicator or Alarm
Components in dashed line are supplied by customer.
Содержание 755A
Страница 1: ...Rosemount Analytical MODEL 755A OXYGEN ANALYZER INSTRUCTION MANUAL 245364 U ...
Страница 16: ...PREFACE P 8 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 34: ...UNPACKING AND INSTALLATION 2 14 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 50: ...ROUTINE OPERATION 4 2 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 76: ...ELECTRONIC CIRCUIT ANALYSIS 6 16 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 86: ...ROUTINE SERVICE AND MAINTENANCE 7 10 April 2000 Rosemount Analytical 245364 U Model 755A Oxygen Analyzer NOTES ...
Страница 90: ... blank ...
Страница 92: ... blank ...
Страница 94: ... blank ...