Rosemount Model 3144 and 3244MV Smart Temperature Transmitters
3-4
TABLE 3-3. Valid Options/Outputs Using Both Sensor 1 and Sensor 2 (Model 3244MV Only).
Primary
Variable
H
o
t
B
ack
u
p
Ena
ble
d
Dri
ft
Al
e
rt
A
c
ti
vat
e
d
Dri
ft
Al
a
rm
M
ode
O
n
Se
ns
or
1
F
a
il
Se
ns
or
2
F
a
il
Analog
Output
Digital
Status
Diff
Temp
Value
Term
Temp
Value
Sensor
1
Value
Sensor
2
Value
Average
Temp
Value
Differential
Y
Y/N
Y/N
Y/N
Y/N
Invalid
Term Temp
Y
Y/N
Y/N
Y/N
Y/N
Invalid
Sensor 2
Y
Y/N
Y/N
Y/N
Y/N
Invalid
Any
N
N
Y/N
N
Y
Alarm
Sensor 2 Fail
±9999
Normal
Normal
±9999
±9999
Any
N
N
Y/N
Y
N
Alarm
Sensor 1 Fail
±9999
Normal
±9999
Normal
±9999
Any
N
N
Y/N
Y
Y
Alarm
Sensor 1/Sensor 2 Fail
±9999
Normal
±9999
±9999
±9999
Any
N
Y
N
N
Y
Alarm
Drift Alert/Sensor 2 Fail
±9999
Normal
Normal
±9999
±9999
Any
N
Y
N
Y
N
Alarm
Drift Alert/Sensor 1 Fail
±9999
Normal
±9999
Normal
±9999
Any
N
Y
N
Y
Y
Alarm
Sensor 1/Sensor 2 Fail
±9999
Normal
±9999
±9999
±9999
Any
N
Y
Y
N
N
Alarm
Drift Alert
Normal
Normal
Normal
Normal
Normal
Any
N
Y
Y
N
Y
Alarm
Drift Alert/Sensor 2 Fail
±9999
Normal
Normal
±9999
±9999
Any
N
Y
Y
Y
N
Alarm
Drift Alert/Sensor 1 Fail
±9999
Normal
±9999
Normal
±9999
Any
N
Y
Y
Y
Y
Alarm
Sensor 1/Sensor 2 Fail
±9999
Normal
±9999
±9999
±9999
Differential
N
N
Y/N
N
N
Differential
None
Normal
Normal
Normal
Normal
Normal
Differential
N
Y
N
N
N
Differential
Drift Alert
Normal
Normal
Normal
Normal
Normal
Term Temp
N
N
Y/N
N
N
Term Temp
None
Normal
Normal
Normal
Normal
Normal
Term Temp
N
Y
N
N
N
Term Temp
Drift Alert
Normal
Normal
Normal
Normal
Normal
Sensor 1
N
N
Y/N
N
N
Sensor 1
None
Normal
Normal
Normal
Normal
Normal
Sensor 1
N
Y
N
N
N
Sensor 1
Drift Alert
Normal
Normal
Normal
Normal
Normal
Sensor 1
Y
N
Y/N
N
N
Sensor 1
None
Normal
Normal
Normal
Normal
Normal
Sensor 1
Y
N
Y/N
N
Y
Sensor 1
Sensor 2 Fail
±9999
Normal
Normal
±9999
Sens 1
Sensor 1
(1)
Y
N
Y/N
Y
N
Sensor 2
Hot BU/Sensor 1 Fail
±9999
Normal
±9999
Normal
Sens 2
Sensor 1
Y
N
Y/N
Y
Y
Alarm
Hot BU/Sensor 1/Sensor
2 Fail
±9999
Normal
±9999
±9999
±9999
Sensor 1
Y
Y
N
N
N
Sensor 1
Drift Alert
Normal
Normal
Normal
Normal
Normal
Sensor 1
Y
Y
N
N
Y
Sensor 1
Drift Alert/Sensor 2 Fail
±9999
Normal
Normal
±9999
Sens 1
Sensor 1
Y
Y
N
Y
N
Sensor 2
Drift Alert/Hot BU/Sensor
1 Fail
±9999
Normal
±9999
Normal
Sens 2
Sensor 1
Y
Y
N
Y
Y
Alarm
Hot BU/Sensor 1/Sensor
2 Fail
±9999
Normal
±9999
±9999
±9999
Any
Y
Y
Y
Y/N
Y/N
Invalid
(2)
Sensor 2
N
N
Y/N
N
N
Sensor 2
None
Normal
Normal
Normal
Normal
Normal
Sensor 2
N
Y
N
N
N
Sensor 2
Drift Alert
Normal
Normal
Normal
Normal
Normal
Average
N
N
Y/N
N
N
Average
None
Normal
Normal
Normal
Normal
Normal
Average
N
Y
N
N
N
Average
Drift Alert
Normal
Normal
Normal
Normal
Normal
Average
Y
N
Y/N
N
N
Average
None
Normal
Normal
Normal
Normal
Normal
Average
Y
N
Y/N
N
Y
Average
Sensor 2 Fail
±9999
Normal
Normal
±9999
Sens 1
Average
Y
N
Y/N
Y
N
Sensor 2
Hot BU/Sensor 1 Fail
±9999
Normal
±9999
Normal
Sens 2
Average
Y
N
Y/N
Y
Y
Alarm
Hot BU/Sensor 1/Sensor
2 Fail
±9999
Normal
±9999
±9999
±9999
Average
Y
Y
N
N
N
Average
Drift Alert
Normal
Normal
Normal
Normal
Normal
Average
Y
Y
N
N
Y
Average
Drift Alert/Sensor 2 Fail
±9999
Normal
Normal
±9999
Sens 1
Average
Y
Y
N
Y
N
Sensor 2
Drift Alert/Hot BU/Sensor
1 Fail
±9999
Normal
±9999
Normal
Sens 2
Average
Y
Y
N
Y
Y
Alarm
Hot BU/Sensor 1/Sensor
2 Fail
±9999
Normal
±9999
±9999
±9999
Any
N
N
Y/N
N
Y
Alarm
Sensor 2 Fail
±9999
Normal
Normal
±9999
±9999
NOTE: If alarm value is set to low, the value will be –9999, and if set to high the value will be +9999.
NOTE: If a hardware error occurs, all outputs will go to ±9999.
(1) Remapping occurs in this situation.
(2) Hot Backup and Drift Alarm mode can not be used simultaneosly.