Manual Alignment and Recording Correction Values
R&S ZVB
1145.1078.82 2.10
E-1
The measurement can be performed simultaneously on all ports using a setup comprising an LO
divider, four power splitters, two NRVDs and four NRV-Z51 sensors (for 2-port models: two power
splitters and one NRVD with two sensors).
Abs b
all
Ports
REF and MEAS channel relative measurement (a and b wave)
Test setup:
Follow the instructions on the screen
(screw SHORT to portx or connect test
cable between two ports)
Measurement:
Wave ratio b1/ a1
Specific measurement results can be displayed as a graph by clicking on the appropriate button in the
Load
section. This means that a check can be made at a later date. Spurious measurements (outliers)
can then be detected.
The ZVAB-Service program automatically writes the correction data to the Eprom.
rel
all
Ports
Содержание ZVB 4