Troubleshooting R&S
ZVB
1145.1078.82 3.44
E-1
Troubleshooting - Power-up Problems
•
Fault: It is not possible to turn on the ZVB.
Action
Cause of fault / remedy
Check mains switch on the rear panel
⇓
Mains switch OFF: Turn on at mains switch.
Check LED is yellow (standby)
⇓
LED does not come on:
¾
Measure
voltage
at X92.C23
(Front module controller):
Nom. value: +13.5 V
±
1V
Nom. value reached: Fault in key pad or controller.
No voltage:
Power supply defective or short to12 V standby.
Turn on instrument. Check LED is green
⇓
LED does not come on:
¾
Measure the PWR-ON signal at power supply
X92.B24:
< 1V for ON
Voltage > 1V: Key pad or controller defective.
•
Fault: Fan not working.
Action
Cause of fault / remedy
Check voltage at connector:
X35, X36, X37 pin 3:
nom. value 12V
⇓
If no voltage can be measured the fan fuse may be
defective (F12, F62, F63)
Select Service Function
Set Service Level 1
Set Service Function 2.5.0.11.1.5
(max. fan speed)
Check voltage at connector:
X35, X36, X37 pin 3:
nom. value 0.9V
Содержание ZVB 4
Страница 46: ...Table of Contents Alignment R S ZVB 1145 1078 82 I 2 2 E 1 This page has been left blank intentionally ...
Страница 57: ...Instrument Construction and Function Description R S ZVB 1145 1078 82 I 3 2 E 1 ...
Страница 133: ...1145 1078 82 5 5 E 1 Spare Parts List Mechanical Drawings ...
Страница 142: ...1145 1078 82 5 15 E 1 Block Circuit Diagram ...
Страница 143: ......
Страница 144: ......
Страница 145: ......
Страница 146: ......
Страница 148: ......
Страница 149: ......
Страница 150: ......
Страница 151: ......
Страница 152: ......
Страница 153: ...1145 1078 82 5 17 E 1 Circuit diagrams ...
Страница 154: ......
Страница 155: ......
Страница 156: ......