Troubleshooting R&S
ZVB
1145.1078.82 3.66
E-1
Testing the Network Controller Board
Testing the IF inputs
It is assumed that there is one functioning reflectometer in the instrument.
¾
Disconnect the IF-MEAS and IF-REF cable from each of the reflectometers.
¾
Connect the input to be tested at the end of the appropriate IF cable (W136, W137, W138, W139,
and W146, W147, W148, W149) to a functioning port using the adapter cable and SMA-N adapter .
¾
Set the ZVB to CW sweep mode, CENTER 20 MHz.
¾
Setting at the port used for the measurement: POWER -10 dBm
¾
Setting at the port associated with the network controller under test: WAVE QUANTITY ax or bx.
¾
Enter service function 2.21.2.18.20000000 (IF = 20 MHz).
¾
Disable level corrections with SF 2.21[..24].2.15.1
If the level displayed on the ZVB's screen is not within the range -4 dBm ± 2 dB, the board must be
replaced.