
Performance Test
R&S
ZVB
1145.1078.82 1.6
E-1
Checking the Accuracy of Output Power
Instrument: Power
sensor
or
Power meter with power sensor
(see Chapter “Test Equipment“, Item2)
Test setup:
Connect the power sensor to port 1 (port 2 to port 4)
Power sensor settings:
For measurement frequencies, see Performance Test Report {f
GEN
*}.
R&S
®
ZVB settings:
- [
Preset
]
- [
Meas
:Wave Quantities: a1 Src Port 1
(a2 Src Port 2, a3 Src Port 3, a4 Src Port 4)]
- [
Sweep
: Sweep Type : CW Mode :
CW Frequency :
{f
GEN
*}
;
Power :
-10 dBm
]
- [
Sweep
: Single :
Restart
]
* For measurement frequencies, see Performance Test Report {f
GEN
}.
Measurement:
Read off the levels indicated by the power sensor
Level deviation = L
SENSOR
– (-10 dBm)
Содержание ZVB 4
Страница 46: ...Table of Contents Alignment R S ZVB 1145 1078 82 I 2 2 E 1 This page has been left blank intentionally ...
Страница 57: ...Instrument Construction and Function Description R S ZVB 1145 1078 82 I 3 2 E 1 ...
Страница 133: ...1145 1078 82 5 5 E 1 Spare Parts List Mechanical Drawings ...
Страница 142: ...1145 1078 82 5 15 E 1 Block Circuit Diagram ...
Страница 143: ......
Страница 144: ......
Страница 145: ......
Страница 146: ......
Страница 148: ......
Страница 149: ......
Страница 150: ......
Страница 151: ......
Страница 152: ......
Страница 153: ...1145 1078 82 5 17 E 1 Circuit diagrams ...
Страница 154: ......
Страница 155: ......
Страница 156: ......