
Performance Test
R&S
ZVB
1145.1078.82 1.14
E-1
Checking the DC Measurement Inputs
Test equipment:
DC Power Supply NGSM 32/10
(see Chapter “Test Equipment“, Item9)
Multimeter URE3
(see Chapter “Test Equipment“, Item10)
DC cable
(see Chapter “Test Equipment“, Item11)
Test setup:
Connect the Power Supply to the Input DC MEAS 1V (DC MEAS 10V) of
the ZVB using the DC cable
R&S
®
ZVB settings:
- [
Preset
]
For DC Meas 1 V:
- [
Meas
: More : DC Inputs : DC Meas
±
1 V]
- [
Format
: Real]
- [
Scale
: Scale/Div : .25 x1]
- [
Marker
]
For DC Meas 10 V:
- [
Meas
: More : DC Inputs : DC Meas
±
10 V]
- [
Format
: Real]
- [
Scale
: Scale/Div : 2.5 x1]
- [
Marker
]
Measurement:
Set Power Supply to
DC values U
DC
using Multimeter URE (DC values
see Performance Test Report) and connect it to pos. and neg. input
Read off the DC level U
DC ZVB
indicated by the marker
.
Calculation:
Deviation = U
DC
-
U
DC ZVB
Содержание ZVB 4
Страница 46: ...Table of Contents Alignment R S ZVB 1145 1078 82 I 2 2 E 1 This page has been left blank intentionally ...
Страница 57: ...Instrument Construction and Function Description R S ZVB 1145 1078 82 I 3 2 E 1 ...
Страница 133: ...1145 1078 82 5 5 E 1 Spare Parts List Mechanical Drawings ...
Страница 142: ...1145 1078 82 5 15 E 1 Block Circuit Diagram ...
Страница 143: ......
Страница 144: ......
Страница 145: ......
Страница 146: ......
Страница 148: ......
Страница 149: ......
Страница 150: ......
Страница 151: ......
Страница 152: ......
Страница 153: ...1145 1078 82 5 17 E 1 Circuit diagrams ...
Страница 154: ......
Страница 155: ......
Страница 156: ......