Instrument functions
R&S
®
LCX Series
82
User Manual 1179.2260.02 ─ 02
a) Select "Short Correction".
b) Turn
"Enabled".
c) Select "Start Short Correction (Spot)".
The R&S
LCX displays a message that prompts you to short-circuit the termi-
nals.
d) Short-circuit the contacts, i.e. insert the supplied short circuit plate into the slots
of the test fixture.
e) Confirm the message with "Yes".
The R&S
LCX displays a progress message until the correction is completed.
The short spot correction takes only a few seconds.
Aligning the instrument using load correction
For executing load correction, the R&S
LCX requires a reference value of any compo-
nent to derive the results of the following measurements.
The component can be a resistor or a capacitor with known impedance. I.e. a compo-
nent that is similar in value and shape to the samples you want to measure, or one of
the series of samples used as default (standard) value.
Prerequisite is also, that you execute the open and short correction before.
1. Execute the short and open spot corrections, as described in
ment on a dedicated working point"
2. Select "Load Correction".
3. Set the resistance, and, the reactance values of the reference component, e.g.:
a) Set "Reference Load R" =
10
k
Ω
b) Set "Reference Load X" =
0
Ω
4. Perform a load spot correction:
a) Select "Load Correction".
b) Turn
"Enabled".
c) Connect the reference component.
d) Select "Start Load Correction (Spot)".
8.7
Configuration preset
The configuration preset function enables you to assign up to three measurement con-
figurations as favorites to the preset keys. By selecting a favorite, the R&S
LCX
changes all predefined parameter settings at once, and thus provides to switch quickly
between different measurement settings and conditions.
About configuration preset
Aspects on the functionality:
●
Included parameters
Configuration preset
www.allice.de