Getting started
R&S
®
LCX Series
33
User Manual 1179.2260.02 ─ 02
a) Press the [settings] key at the front panel.
On the front panel, press the [settings] button.
b) In the "Device" tab, select "Save/Recall Device Settings".
c) Select "Default Settings".
The instrument prompts you to confirm the operation.
d) Confirm with "Ok".
The R&S
LCX resets all parameters except for system settings.
e) Select to return to the settings menu.
Connecting the test fixture
The figure shows schematically the elements to be connected.
► To connect the test fixture, proceed as described in
Setting up the test signal
To configure the test signal:
1. To set the frequency:
a) On the screen, select "Frequency".
b) Enter, e.g. 10
kHz using the keypad on the screen.
c) Confirm with
.
2. To select the level parameter:
Before setting the test signal level, you can select either the signal voltage, or the
signal current for the level setting:
a) Press the [settings] at the front panel.
b) Select the "Measurement" tab.
c) Select "Test Signal Level".
d) Select, e.g. "Voltage".
e) Confirm with "Set".
f) Press the [home] key to return to the measurement window.
The R&S
LCX provides the selected parameter for the level setting on the screen.
3. To set the test signal level:
a) On the screen, select "Level".
b) Enter, e.g. 1.3
VRMS.
c) Confirm with
.
4. If you know the expected range, set the range as follows.
Otherwise skip this step and proceed with
a) Select "Range".
b) In the "Select Measurement Range" dialog, select the expected impedance,
e.g. 10
k
Ω
.
c) Confirm with "Set".