Instrument functions
R&S
®
LCX Series
101
User Manual 1179.2260.02 ─ 02
"Step Size"
Uses the interval between one sweep parameter value to the next,
set with
"Number of Points"
Uses the number of values within the range of minimum and maxi-
mum sweep parameter settings, set with
Remote command:
Step Size
Sets the interval between each sweep step within the measurement range
).
Depending on the selected sweep parameter, the R&S
LCX provides the correspond-
ing value and unit.
"xx V"
For
Sweep Parameter > Test Signal Voltage or Voltage Bias
step width in V.
"xx Hz"
For
, sets the time interval in Hz.
"xx A"
For
Sweep Parameter > Current Bias
, sets the step width in A.
Remote command:
Number of Points
Sets the number of measurement points within the measurement range (
).
Remote command:
Abort
Cancels a dynamic impedance measurement process while it is running.
Remote command:
8.8.4.2
Using the dynamic impedance measurement function
The dynamic impedance measurement function records the measurement readings
automatically. The R&S
LCX stores the data in a log file with a generically created file-
name, either on the USB stick or in the instrument internal memory directory.
Selecting the memory location
To determine the destination for saving the data file:
► Follow the steps described in
"Selecting the memory location"
Measuring in dynamic impedance measurement mode
1. Press the [settings] key at the front panel.
2. In the "Functions" tab, select "Dynamic Impedance Measurement".
3. Configure the required settings.