Maintenance, storage and disposal
R&S
®
LCX Series
225
User Manual 1179.2260.02 ─ 02
14
Maintenance, storage and disposal
The product does not require regular maintenance. It only requires occasional clean-
ing. It is however advisable to check the nominal data from time to time.
14.1
Cleaning
How to clean the product is described in
Do not use any liquids for cleaning. Cleaning agents, solvents (thinners, acetone),
acids and bases can damage the front panel labeling, plastic parts and display.
14.2
Changing fuses
The R&S
LCX is protected by two fuses, you can check and replace, if there are any
problems:
●
F1: Line fuse
If the R&S
LCX does not start, it is possible that a blown line fuse is the cause. In
this case, check the line fuse and replace it, see
.
●
F2: Bias fuse
Protects the bias voltage input from overload to avoid damage of the circuitry. Reg-
ularly check the condition of the fuse and replace it in case it is blown or damaged,
see
"Replacing the external bias fuse"
If you have still problems, contact Rohde
&
Schwarz customer service at
14.3
Storage
Protect the product against dust. Ensure that the environmental conditions, e.g. tem-
perature range and climatic load, meet the values specified in the data sheet.
14.4
Disposal
Rohde
&
Schwarz is committed to making careful, ecologically sound use of natural
resources and minimizing the environmental footprint of our products. Help us by dis-
posing of waste in a way that causes minimum environmental impact.
Disposal
www.allice.de
Allice Messtechnik GmbH