Troubleshooting
R&S
®
LCX Series
222
User Manual 1179.2260.02 ─ 02
12
Troubleshooting
................................................................................ 222
Problems during firmware update
......................................................................... 222
Cannot establish a LAN connection
......................................................................223
................................................................................ 223
12.1
Displaying status information
The R&S
LCX displays status information in diofferent ways:
●
Status bars
Using icons, the status bars indicate general instrument configuration and operat-
ing modes, and measurement modes and states, see
●
Messages
Messages that appear on the screen inform you on current settings and states.
The R&S
LCX distinguishes between status messages, information messages, and
also warning and error messages.
Some information and messages are entered in the error/event queue of the status
reporting system, see
Chapter 10.2.4, "Status reporting system"
12.2
Problems during firmware update
The firmware update is described in
tions for potential problems that can occur during firmware update are described in the
following sections.
Firmware update was interrupted
If, for example, a power cut happened during the firmware update, problems can occur.
1. Perform the firmware update again. Sometimes, a further update fixes the prob-
lems.
2. If you nevertheless can not access the R&S
LCX, contact the local service repre-
sentative, see
Chapter 12.4, "Contacting customer support"
Firmware update was aborted
If there is not enough free memory space, the firmware update aborts. The instrument
displays an error message.
1. Execute the sanitzation procedure, as described in the instrument security proce-
dures. You find this document on the product page at:
2. Start the firmware update again.
Problems during firmware update