Instrument functions
R&S
®
LCX Series
78
User Manual 1179.2260.02 ─ 02
Impedance
Equivalent circuit
0
≤
Z
≤
S
serial measurement
S
<
Z
<
range
parallel measurement
8.6.2
Open/short/load correction settings
The correction dialog enables you to activate and execute the correction techniques
provided by the instrument.
Depending on the connected test fixture and the type of component sample to be mea-
sured, refer to the description on how to execute the corresponding corrections, see
Chapter 6.1, "About test fixtures"
Access:
1. Press the [settings] key at the front panel.
2. Select the "Measurement" tab.
3. Select "Open/Short/Load Correction".
Figure 8-3: Open/Short/Load correction dialog
In this menu, you can activate the corresponding corrections directly, and access
the subdialogs to configure and execute the corrections.
The required remote commands are described in:
Chapter 11.6, "CORRection sub-
How to: see
Chapter 8.6.3, "Setting measurement parameters"
If you need to perform the open and short correction frequently, you can assign these
functions to the [
★
(User)] key and execute them with just one keystroke, see
Settings
Open/Short/Load Correction Enabled
........................................................................... 79
Start Short/Open Correction (Spot)
...............................................................................79
Start Short/Open Correction (Full)
................................................................................ 79
Measurement parameters
www.allice.de
Allice Messtechnik GmbH