Network operation and remote control
R&S
®
LCX Series
163
User Manual 1179.2260.02 ─ 02
The instrument must reboot to load the corresponding protocol driver.
5. To proceed, you have the following options:
a) Confirm with "OK".
The R&S
LCX reboots and activates the connection. It indicates the selected
USB interface port in the device status bar on top of the screen.
b) If you want to undo your selection, tap any point of the screen outside the mes-
sage.
The message disappears and you can correct your selection.
Assigning the GPIB address manually
Option: R&S
NG-B105 GPIB (IEC 625/IEE 488) bus interface
The controller addresses the instrument within the GPIB bus channel, providing
addresses from 0 to 30. The R&S
LCX has the factory-set address 20, but you can
adjust it manually according to your network environment. It is assumed, that you have
established the connection, see
"Establishing the GPIB remote control connection"
To select and enter the GPIB manually:
1. To access the setting, the [settings] key at the front panel.
2. In the device settings menu, select "Interfaces" > "GPIB Address".
An on-screen keypad opens.
3. Enter the channel number.
4. Confirm with
.
The R&S
LCX changes the GPIB address accordingly.
Note:
The address is maintained after a reset of the instrument settings.
10.6
Operating the R&S
LCX remotely
When you switch on the instrument, it is always in manual operation state (local state).
You can operate it using the controls on the front panel. To operate the instrument from
a remote location, see the following possibilities, how to set up remote control connec-
tions over the available interfaces.
Prerequisites to proceed with the following instructions are:
●
You have set up the connection between the R&S
LCX and the controller, see
Chapter 10.4, "Connecting the R&S
●
The instrument and the controller are turned on and ready for operation.
Operating the R&S
LCX remotely
www.allice.de