Troubleshooting and Repair
R&S
®
ATS1000
72
Instructions Handbook 1179.2983.02 ─ 01
10. Reattach the acrylic glass window.
11. Secure the acrylic glass window with the four Torx
10 screws.
To restore power to the chamber
If the chamber unexpectedly remains inactive, proceed as follows:
1. Check the connection to mains power.
2. If the chamber is disconnected from mains power, reconnect it.
3. If the chamber still has no power, check its automatic fuse (
4. If the fuse has been triggered, check if any movable parts inside the chamber are
blocked mechanically, as described in
"To enable the automatic fuse"
5. If no movable parts are blocked, enable the fuse as described in
6. If the chamber still has no power, check the voltage of the mains power.
7. If the mains power is not live, switch it on again.
8. If the chamber still has no power, switch the fuse off and on again, as described in
"To enable the automatic fuse"
9. If the chamber still has no power, contact Rohde
&
Schwarz
To restore the door function
If you cannot open or close the door properly, proceed as follows:
1. Press the [Laser] button on the front panel so that the button's LED lights up.
2. Disconnect the mains power.
3. With the [Laser] button in its "on" position, wait until the button's LED goes out.
Waiting until the LED goes out ensures that capacitors inside the chamber's lower
compartment are no longer charged.
4. Reconnect the mains power.
5. Check the door function.
The chamber performs an automatic referencing procedure for the door lock; see
Chapter 7.3.2, "Referencing the Door's Lock"
6. If the door still does not open or close properly, contact Rohde
&
.
The storage and operating temperature ranges for the R&S
ATS1000 are specified in
the data sheet.
Chamber Troubleshooting