Installation and Commissioning
R&S
®
ATS1000
38
Instructions Handbook 1179.2983.02 ─ 01
The wooden wedges of the front panel rest on the floor so that the front panel
forms a ramp for the chamber.
10. On the outside of the crate, unscrew the timber beams that prevent the chamber
from moving out of the crate.
11. Remove the timber beams.
12. If accessories are included with the chamber, take the accessories out of the crate.
13. Release the brakes of the front wheels. See
.
14.
WARNING!
The chamber is heavy. Wear protective clothing, especially safety foot-
wear with protective toe caps.
Move the chamber out of the crate and down the ramp. See
When moving the chamber, hold it at any solid parts of its walls.
15. Engage the wheel brakes.
16. Remove the antistatic wrap from the chamber.
17. Keep the original packing material. Use it if you need to transport the chamber
later.
We recommend storing the transport crate in its original upright position so that the
tilt indicators can be used again for future transportation.
18. Check the delivery against the delivery notes or accessory list to ensure that all
items are present.
19. Check the chamber for damage.
If the delivery is incomplete or equipment is damaged, contact Rohde
&
Schwarz.
6.3
Installing the Chamber
Familiarize yourself with residual risks and potentially dangerous situations.
See
Chapter 2.3, "Potentially Danger-
The chamber does not require any special installation. Instead:
1. Move the chamber to its operating site.
Chapter 6.1, "Choosing the Operating Site"
Chapter 5.1, "Moving the Chamber"
2. Position it for optimum accessibility, for example next to a rack holding your other
test equipment.
Installing the Chamber