Machine Overview
R&S
®
ATS1000
21
Instructions Handbook 1179.2983.02 ─ 01
chamber is configured at the factory. Feedthroughs must, therefore, always be installed
by the manufacturer only. If you want to exchange or add feedthroughs, contact
Rohde
&
or your sales representative. The available feedthrough
Only an
is allowed to connect, disconnect or exchange cables at the feed-
throughs.
Only an
is allowed to remove, mount or open the laser boxes (3 and 4), to
align the lasers, and to use the lasers for aligning the positioner axes.
Only Rohde
&
are allowed to open the service panel (5).
Feedthroughs on the left panel
The following feedthroughs are installed by default on the left panel of the chamber:
Figure 4-4: Feedthroughs on the chamber's left panel (left when viewed from the front)
Left = Twin RF feedthrough [A111] (= top row), with measurement cables connected for vertical [VER] and
horizontal [HOR] polarization
Right = 8 factory-configurable feedthroughs on the lower left panel (= middle and lower row)
Table 4-1: Feedthroughs on the left panel
Position
1st column
2nd column
3rd column
4th column
Top row
[A111], reserved for the two RF ports that feed RF test signals through the rotational axis of
the antenna elevation positioner arm
Middle row [A121]
Standard config-
uration:
LAN feedthrough
[A122]
Standard config-
uration:
USB feedthrough
[A123]
Standard configu-
ration:
D-Sub feedthrough
[A124], standard configuration:
Right: not connected
Left: to DUT on azimuth turnta-
ble, feedthrough depends on
selected RF cable set, either for
40
GHz or for 50
GHz.
●
For 40
GHz: 2.92
mm RF
feedthrough
●
For 50
GHz: 1.85
mm RF
feedthrough
Lower row
[A131]
Standard config-
uration:
Not connected
[A132]
Standard config-
uration:
Not connected
[A133]
Standard configu-
ration:
Not connected
[A134]
Standard configuration:
Not connected