Safety
R&S
®
ATS1000
14
Instructions Handbook 1179.2983.02 ─ 01
For detailed instructions, see
Chapter 6, "Installation and Commissioning"
Connecting to power
The chamber is an overvoltage category II product. Connect it to a fixed installation
used to supply energy-consuming equipment such as household appliances and simi-
lar loads. Bear in mind that electrically powered products pose risks, such as electric
shock, fire and potentially fatal injury.
Take the following measures for your safety:
●
Before connecting the chamber to a power source (your mains), ensure that this
source matches the voltage and frequency range indicated in the data sheet.
●
Only use the power cable that is delivered with the chamber. It complies with coun-
try-specific safety requirements.
●
Only insert the power cable's plug into an outlet with protective conductor terminal.
●
Only use intact cables and route them carefully so that they cannot be damaged.
Check the power cables regularly to ensure that they are undamaged.
●
Only connect the chamber to a power source with a fuse protection of maximum
20
A.
●
Ensure that you can disconnect the power supply unit from the power source at
any time. Pull the power plug to disconnect the chamber. The power plug must be
easily accessible.
●
Install an easily accessible panic button (power-off switch, not included in the deliv-
ery) for interrupting the power source to the chamber.
Operating the door
The risks when handling the door are described in
"Door of the chamber is heavy"
"Risk of crushing fingers when moving the door"
Open and close the door by its handle. Do not touch other parts of the door. When you
push the [LOCK / UNLOCK] button, the automatic door locking mechanism is activa-
ted. It pulls the door tightly against the chamber with great force in order to ensure
effective sealing.
Establish safety rules for operating the door to ensure that nobody gets their fingers
caught between the door and the body of the chamber.
Use the chamber as configured by Rohde
&
Schwarz. Never tamper with its safety
installations.
For detailed instructions, see
Chapter 7.3, "Operating the Door"
Maintenance
Carry out the maintenance tasks as required, to ensure that the chamber continues to
function properly and maintain the safety of all users. For detailed instructions, see
Chapter 8, "Inspection and Maintenance"
Potentially Dangerous Situations