Getting Started
R&S
®
OSP
27
User Manual 1178.7100.02 ─ 05
For details on mounting limitations, refer to
Chapter 2.4.1, "Module Integration Over-
2.2.1.4
Front Panel of the R&S
OSP-B200S2 Satellite Unit
For a comprehensive description of this unit, also refer to the
R&S
OSP-B200R/
B200S2 Satellite System user manual
, available for download at
The R&S
OSP-B200S2 is a standalone device, only, not to be mounted into a 19" rack.
The front panel of satellite unit the features 2 module slots, into which you can insert 1-
slot or 2-slot modules:
Figure 2-6: Front view of the R&S
OSP-B200S2 (2HU)
SlotA = Front slot A, here with a 1-slot switch module
SlotB = Front slot B, here with a 1-slot switch module
This unit is designed to serve as a satellite with up to 2 switch modules, controlled from
a base switch unit with remote control module
.
For example, you can use the satellite unit inside a shielded chamber for EMC tests
that do not tolerate electrical wiring. In this scenario, use the satellite with local battery
power supply, and control the switch modules fitted in the unit via a fiber-optic link. See
also
"Wired Link versus Fiber-Optic Link"
Instrument Tour