8. Enter the host name of your primary switch unit into the address bar of your
browser.
For example, enter
osp320-101357
(without additional characters).
Note that you can also enter a switch unit's IP address into the browser's address
bar, instead of the host name.
The browser shows the primary device's graphical user interface.
9. Go to the
dialog.
10. Click "Add a secondary device".
11. Enter the host name of a secondary switch unit into the "Address" field.
Note that you can also enter a switch unit's IP address into the "Address" field, but
we recommend using the host name.
12. Confirm your entry.
13. Repeat
for each secondary switch unit.
You can use all connected switch units via the
If there are problems with this procedure, refer to
4.4
Operation in Interconnection Mode
Access:
Complex switching scenarios can require several switch units with many modules
working together. The R&S
OSP allows connecting and jointly controlling even many
switch units.
As a precondition, all connected switch units must run the same firmware version.
However, controlling this combination of devices can be a challenging task, if each
switch unit must be addressed separately.
Controlling interconnected setups in compatibility mode is limited to 1 primary switch
unit with frame ID "F01" and 8 secondary switch units with frame IDs "F02" to "F09".
To facilitate the combination of switch units, the R&S
OSP features operation in "Inter-
connection" mode. To achieve this operating mode, define one of your devices as a pri-
mary device and all other as secondary devices:
Operation in Interconnection Mode