RC Commands
R&S
®
OSP
287
User Manual 1178.7100.02 ─ 05
Manual operation:
See
TRIGger:SEQuence:DEFine:LENGth
<integer>
Sets or queries the length of the sequence for the "Sequenced" trigger.
Parameters:
<integer>
Range:
3 to 16
Example:
Setting:
TRIG:SEQ:DEF:LENG 5
Query:
TRIG:SEQ:DEF:LENG?
Reply:
5
Manual operation:
See
TRIGger:SEQuence:DEFine[:ENTRy]
<integer>,<string>
Sets or queries the path, which is defined for the selected number in the sequence.
Parameters:
<integer>
Selects a number in the trigger sequence.
Range:
0 to (LENGth - 1), see "TRIG-
ger:SEQuence:DEFine:LENGth"
<string>
Specifies the path name to be set for the selected number in the
trigger sequence. Write the path name in quotation marks.
Example:
Setting:
TRIG:SEQ:DEF 1,"PathB"
Query:
TRIG:SEQ:DEF:ENTR? 1
Reply:
"PathB"
Manual operation:
See
TRIGger:SEQuence:DEFine:ALL
<string>
Sets several or all paths for the sequenced trigger. The query returns all path names.
Parameters:
<string>
Specifies all path names for the trigger sequence. Separate the
path names by commas and write each path name in quotation
marks.
Example:
Setting:
TRIG:SEQ:DEF:ALL "PathA","PathB","PathC",
"PathD","PathE"
Query:
TRIG:SEQ:DEF:ALL?
Reply:
"PathA","PathB","PathC","PathD","PathE"
TRIGger:SIGNal:LEVel
<numeric_value>
Sets or queries the voltage level, which defines the threshold for a trigger event. The
level setting is not influenced by reset commands.
TRIGger Commands