Calibration
Power Supply NGMO
The display must show the following:
Enter new CAL. Date:
YY/MM/DD: yy/mm/dd
•
yy is the current year (from 01 to 99).
•
mm is the current month (from 01 to 12)
•
dd is the current day (from 01 to 31 –
depends on the selected month):
25. Enter the Date.
Use the variation knob and the arrow keys to enter the current
date.
The display must show the following:
Calibration success-
fully completed!
The calibration procedure must be repeated for the other
channel (NGMO2 only). This can be done in the following way.
26. Connect calibration adapter to output terminal B.
27. Press the “CHN A/B/DVM” key once. (Channel B).
28. Select Menu Item B38.
29. Repeat all the other steps listed above.
3.00 / 05-2014