Power Supply NGMO
Calibration
11. Enter the Measuring value.
Use either the variation knob or numerical entry keys (press shift
key to initiate the numerical input) to enter the measured value
that is displayed on the DVM.
12. Press ENTER.
The display must show the following:
Look at DVM and ent.
Read value: xx.yyy V
13. Enter the Measuring value.
Use either the variation knob or numerical entry keys to enter the
measured value that is displayed on the DVM.
14. Press ENTER.
NOTE:
When using the calibration set-up specified in section 7.2, the
following display may be neglected.
The display must show the following:
Connect known 1 Ohm
Resistor to out.term
15. Press ENTER.
The display must show the following:
Enter exact Resistor
value: x.xxxx Ohms
NOTE:
The values of the resistors installed in the calibration adapter
can be taken from the calibration certificate supplied with the
calibration adapter.
7-6
3.00 / 05-2014
Содержание NGMO1
Страница 1: ...Service Manual Power Supply NGMO1 and Power Supply NGMO2 192 1500 21 and 192 1500 24 5707 2571 82 ...
Страница 12: ...Power Supply NGMO User Information 1 4 3 00 05 2014 ...
Страница 14: ...Power Supply NGMO Customer Service 2 2 3 00 05 2014 ...
Страница 16: ...Power Supply NGMO Safety 3 2 3 00 05 2014 ...
Страница 21: ...Unit Description Power Supply NGMO Fig 4 2 Power Supply NGMO Function block diagram 3 00 05 2014 4 5 ...
Страница 22: ...Power Supply NGMO Unit Description 4 6 3 00 05 2014 ...
Страница 42: ...Power Supply NGMO Calibration 7 12 3 00 05 2014 ...
Страница 72: ...Power Supply NGMO Interfaces 10 4 3 00 05 2014 ...
Страница 74: ...Power Supply NGMO Bill of Materials 11 2 3 00 05 2014 ...