Subject to change without notice

V e r t i c a l  s y s t e m          

STOP

  The passive probes must be adjusted to the inputs 

to which they are connected. See the probe manual 
for the adjustment procedure. The PROBE ADJUST 
output is only usable for 1:1 and 10:1 probes, for 
100:1 or 1000:1 probes special external genera-
tors with a perfect step response have to be used. 
Please use the shortest possible ground connec-
tion.

The coupling is selected in the short menu: by just pushing the 
appropriate key the coupling is chosen, also the signal may be 
inverted. The menu is valid for the activated channel as indicated 
by the channel key light up. The channel number will be shown 
in the top of the menu. By pushing the respective key of another 
channel the menu will transfer to this channel.

4.2.  Sensitivity, Y-Positioning, and Offset

The sensitivity of the analog inputs can be selected with the large 
knob in the CHANNEL CONTROL section of the front panel in 
1-2-5 steps from 1 mV/div to 5 V/div independent of the 50 Ω or 
1 MΩ selection. The knob is associated with the channel selected 
by pushing the respective key. The sensitivity can be changed 
to continuous control by pushing the knob once. The smaller 
one of the knobs is used for vertical positioning. 

By pushing the MENU key the extended menu is called. On page 
2 of this menu a DC offset can be added to the signal. In order 
to switch this offset in the respective soft key must be pushed. 
The settings window will be backlit in blue, and the activity 
indicator next to the general knob will light up; the offset can 
now be adjusted with the knob. The offset voltage will be added 
to the signal at the vertical amplifi er input offsetting it by that 
amount from the zero position. The possible amount of offset 
depends on the Volts/div setting chosen.  The offset function 
being activated will be indicated by two channel markers on 
the left of the display, also visible if the menu was closed. One 
marker indicates the position, the other the offset (refer to Fig. 
4.3). The offset is individually adjustable for each channel.

Each analog channel may also be shifted in time by ±15 ns. This 
adjustment is selected in the same menu and according to the 
same method as the DC offset; it is used for compensating for 
the different signal delays of voltage and current probes and 
different cable lengths. 

4.3.  Bandwidth Limit and Signal Inversion

An analog 20 MHz low pass can be inserted in the signal path 
in either the short or extended menu. This will eliminate all 

higher frequency interference. The fi lter is activated in the 
short menu by pushing the respective soft key; the information 
fi eld will be backlit in blue, 

BW

 will be displayed in the channel 

information window.

Signal inversion is available in the short and the extended me-
nus. If it is activated the information fi eld will be backlit in blue, 
and there will be a bar above the channel number.

4.4. Probe 

attenuation 

selection

The HZ350 probes supplied with the instrument will be re-
cognized by the instrument which automatically selects the 
appropriate factor. If any other probe without automatic recog-
nition of the attenuation ratio or just a cable is connected to the 
instrument, the attenuation factor can be set manually in the 
extended menu. This is possible for x 1, x 10, x 100 x 1000 or as 
defi ned by the user from x 0.001 to x 1000. 

Fig. 4.3: Vertical offset in the extended menu 

Содержание Hameg HMO3522

Страница 1: ...350 MHz Digital Oscilloscope with FFT and mixed signal option HMO3522 3524 Manual English...

Страница 2: ...different For measurementoperationthefollowinghintsandconditionsregardingemission and immunity should be observed 1 Data cables For the connection between instrument interfaces and external devices c...

Страница 3: ...1 Coupling 15 4 2 Sensitivity Y Positioning and Offset 16 4 3 Bandwidth Limit and Signal Inversion 16 4 4 Probe attenuation selection 16 5 Horizontal System Time Base 17 5 1 Capturing modes RUN and ST...

Страница 4: ...mV 5V div into 1M 50 Offset control 0 2 20V 12div x axis display range Trigger modes slope video pulsewidth logic delayed event FFT for spectral analysis 6 digit counter Autoset automeasurement formul...

Страница 5: ...tenuation ID HZ350 Dual Interface USB RS 232 HO720 CD Optional accessories HO730 Dual Interface Ethernet USB HO740 Interface IEEE 488 GPIB galvanically isolated HZ46 4RU 19 Rackmount Kit HZ355 Slimlin...

Страница 6: ...f visible damage in case loose parts were noticed if it does not function any more after prolonged storage under unfavourable conditions e g like in the open or in moist atmosphere after any improper...

Страница 7: ...ring circuits are considered not connected to the mains if a suitable isolation transformer ful lling safety class II is used Measurements on the mains are also possible if suitable probes like curren...

Страница 8: ...interface STOP Please note the connectors for the active logic probes HO3508 51 52 are solely for theses probes Connecting anything else could destroy the inputs 2 2 Control panel The controls on the...

Страница 9: ...nel actually being displayed these may be selected using the soft keys 3 Below the graticule measurement results of parameters and cursors the settings of the activated vertical channels of the refere...

Страница 10: ...e the rst one will follow 2 7 Basic setting and integrated help Basic settings like language for user interface and help mis cellaneous settings and interface settings can be set using the menu which...

Страница 11: ...r the date and build information Fig 2 11 Updating menu and information window Now choose which to update the rmware or the help function If both are to be updated it is recommended to rst update the...

Страница 12: ...t ground connection possible to the PROBE ADJUST output Connect the ground cable to the probe and the tip to the top of the probe Now connect the compensation box BNC connector to the input connector...

Страница 13: ...ner of the screen In this case the V cursor has se lected the voltages at the two cursor positions their difference and the time difference between the positions will be shown The cursors will be swit...

Страница 14: ...ne or two prede ned mathematical functions A quick setting of mathematical functions is possible by selecting the menu item at the bottom This mode allows you to select the addition or subtraction of...

Страница 15: ...4 1 Front panel area with vertical system controls SCALE VOLTS DIV Y POSITION CHANNEL CONTROL FINE 62 K304 1 1 5 0 1 0 MENU CH1 CH2 CH3 POD1 MATH REF XY CH4 POD2 18 B 19 20 22 23 24 21 25 26 27 By pu...

Страница 16: ...ob The offset voltage will be added to the signal at the vertical ampli er input offsetting it by that amount from the zero position The possible amount of offset depends on the Volts div setting chos...

Страница 17: ...envelope of the signal will be displayed Average Functions only with repetitive signals The universal knob in the CURSOR MENU section of the front panel is used to set the number of signal periods for...

Страница 18: ...tly used functions TYPE selects the type of trigger SLOPE PULSE LOGIC VIDEO and the B TRIGGER SLOPE selects the slope polarity SOURCE opens the menu for the selection of the trigger source FILTER open...

Страница 19: ...he trigger such that rst condition A must be met and then another condition B before the trigger will respond refer to Fig 6 3 E g it is possible to de ne a source channel and a level of 120mV on the...

Страница 20: ...me available which can be shifted up and down with the universal knob A detailed description will follow in the next chapter There are 3 more menu items on the rst page DOTS ONLY The respective soft m...

Страница 21: ...ion to the time elapsed This mode can be selected in the soft menu which will open upon pushing the key INTENS PERSIST the signal intensity can be changed also in this menu Fig 7 2 Menu for setting th...

Страница 22: ...will terminate the XY function if it was active If no or another menu should be active it will be necessary to push the XY key twice for terminating the function 8 Measurements There are two different...

Страница 23: ...SORMODEopensthemenuitemGLUE TO Thismode can be turned on or off If activated the cursors will glue to the signal i e they will automatically follow all changes of the position and scaling controls and...

Страница 24: ...mode offers the following 5 parameters which are directly displayed in the signal Maximum voltage Mean voltage Minimum voltage Rise time Fall time 4 additional parameters will be displayed in the righ...

Страница 25: ...the associated menu In Fig 9 4 in formula MA1 a current of 100 A is added to the channel 1 In the menu for entering constants it is possible to choose from the following list of constants executed by...

Страница 26: ...beshown top left the information about the Zoom and position between both grids and the information about the FFT display span and center frequency is shown below the larger area One of these displays...

Страница 27: ...menu key next to STORE In order to recall stored instrument settings call the main instrument settings menu and select LOAD by pushing the respective soft menu key The data manager will open use the m...

Страница 28: ...addition to references the pure A D converter data can be stored however only on external USB sticks not internally The following formats are available Binary format A binary data set may contain byte...

Страница 29: ...item is called FORMULARIES selection of this menu item will call a submenu which allows you to move sets of for mulas between the internal memory and the USB stick as well as to import or export such...

Страница 30: ...menu eld below by the respective soft key These 6 criteria are available ti t The duration of the bit pattern which will generate the trigger is unequal to the reference time ti t The duration of the...

Страница 31: ...his method all channels may be individually positioned and sized If POD was activated and if the MENU key in the Channel Control section of the front panel was pushed the menu for setting the trigger...

Страница 32: ...e 15 Fig 4 1 Front panel area with vertical system controls 15 Fig 4 2 Short menu for the vertical settings 15 Fig 4 3 Vertical offset in the extended menu 16 Fig 5 1 Control panel of the horizontal s...

Страница 33: ...33 Subject to change without notice A p p e n d i x...

Страница 34: ...34 Subject to change without notice Appendix...

Страница 35: ...35 Subject to change without notice A p p e n d i x...

Страница 36: ...tra e 6 HAMEG Instruments GmbH D 63533 Mainhausen A Rohde Schwarz Company Tel 49 0 61 82 800 0 DQS Certi cation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg com Oscilloscop...

Отзывы: