4.2 Programming a Standard Bolus
There are 2 ways to program a Standard Bolus:
Quick Bolus
Using
a
and
s
on your insulin
pump, with bolus increments
determined by the user
Using Standard
Bolus menu
Using
d
and
f
to enter the menu
and then
a
and
s
to program the
amount
The bolus amount per delivery is limited to 50 units using U100
insulin.
After you program a Standard Bolus, there is a delay of 5 seconds
before bolus delivery starts. This delay allows you to cancel the
bolus by pressing the
a
or
s
key. The Warning W8: BOLUS
CANCELLED is displayed. Press
f
twice to confirm and turn off
the alert. See chapter
8
Warning W8: BOLUS CANCELLED
.
Note
You can not program a new Standard Bolus as long as
one Standard Bolus, Quick Bolus or the immediate part of
a Multiwave Bolus is running.
man_05957613001_E_en.indb 78