Prime the infusion set.
j
Reconnect the infusion set at your infusion site.
j
Put your pump into RUN mode.
j
If the Error E10: CARTRIDGE ERROR occurs during the
rewinding of the piston rod, insert a new battery in the pump
and repeat the procedure.
Error E11: SET NOT PRIMED
The cartridge has been changed but the infusion set has not
been primed. The Error will be displayed when you try to put your
pump in RUN mode.
j
Press
f
twice to snooze and to confirm the Error.
j
Remove or disconnect the infusion set from your infusion site.
j
Prime the infusion set (See section
2.8
Priming the infusion set
).
j
Reconnect the infusion set at your infusion site.
j
Put your pump in RUN mode.
See sections
3.5
Changing the cartridge and infusion set
and
2.8
Priming the infusion set
.
w
WARNING
Never prime an infusion set that is connected to your body.
You risk uncontrolled insulin delivery into your body. Always
follow the instructions provided with your infusion set.