145
7
7.2 Damage to your pump
7.2.1 What to do if you drop your pump
If the pump is dropped it could be damaged, as well as the seal
that protects it against water. So be careful not to drop your
pump! Use the specially designed pump carrying system to help
avoid dropping your pump. See Roche sterile products and
accessories catalogues and brochures for more information, or
contact your local pump support (details on inside back cover).
If you drop your pump:
j
Check that all the connections of your infusion set are still
tight, and if they are not, reconnect or tighten them
j
Check your pump, its sterile products, and accessories for
chips and cracks
j
Change the cartridge if it is damaged
Put your pump in STOP mode, and take out and reinsert the
battery, so that the self-test is started.
Note
If a Temporary Basal Rate, an Extended Bolus or a
Multiwave Bolus is active at the same time, it is cancelled
when your pump is put into STOP mode. A Warning W8:
BOLUS CANCELLED or a Warning W6: TBR CANCELLED is
displayed.
Press
f
twice to confirm and turn off the alert. The
alerts are recorded in the alarm history.
See chapter
8 Warning W6: TBR CANCELLED and
Warning W8: BOLUS CANCELLED.
Make sure that the cancellation was deliberate and then
program a new Temporary Basal Rate (and / or an
Extended or Multiwave Bolus), if needed.
man_05957613001_E_en.indb 145
02.03.2015 10:44:47
Содержание ACCU-CHEK Spirit Combo
Страница 8: ...man_05957613001_E_en indb 6 02 03 2015 10 43 53...
Страница 28: ...24 man_05957613001_E_en indb 24 02 03 2015 10 43 55...
Страница 64: ...60 man_05957613001_E_en indb 60 02 03 2015 10 44 15...
Страница 136: ...132 man_05957613001_E_en indb 132 02 03 2015 10 44 45...
Страница 146: ...142 man_05957613001_E_en indb 142 02 03 2015 10 44 47...
Страница 226: ...222 My notes man_05957613001_E_en indb 222 02 03 2015 10 45 07...