General Technical Data
Maximum time for an Error
E4: OCCLUSION**
Plastic cartridges:
j
at medium basal rate 1.0 U / h: ≤ 2 h
j
at minimum basal rate 0.05 U / h: ≤ 24 h
Aventis Insuman Infusat glass cartridges***:
j
at medium basal rate 1.0 U / h: ≤ 8 h
j
at minimum basal rate 0.05 U / h: ≤ 120 h
Maximum volume before Error
E4: OCCLUSION**
Plastic cartridges: ≤ 1.5 U
Aventis Insuman Infusat glass cartridges***: ≤ 5.0 U
Maximum pressure
400 kPa (4.0 bar), for plastic and Aventis Insuman Infusat glass cartridges*** and infusion
sets.
Flow (delivery rate)
During infusion set priming and bolus: 0.2 U / sec.
Maximum quantity delivered at
a single fault condition
≤ 1.0 U
Cartridge
Accu-Chek Spirit 3.15 ml Cartridge System and Accu-Chek 3.15 ml Plastic Cartridge with a
luer connection. Aventis Insuman Infusat cartridges*** have been evaluated and approved
for use with the Accu-Chek Spirit Combo insulin pump.