46
Hold your pump upright. Hold the new
fi lled cartridge so that the connected
adapter and infusion set points upwards,
parallel, and close to the cartridge
compartment.
Make sure that the top edge of the
threaded part of the adapter is level with
the top of the cartridge compartment.
10.
Press
a
or
s
to move the piston rod
forward until the end plate of the piston
rod is level with the bottom of the
cartridge plunger.
The CARTRIDGE VOLUME decreases
accordingly on the display.
Tip:
To prevent air bubbles move the
piston rod to at least 310 Units.
12.
11.
9.
see step 10
see step 11
man_05957613001_E_en.indb 46
02.03.2015 10:44:07