152
8.1 Confirming a Warning or Error
When you confirm:
W1: CARTRIDGE LOW
E1: CARTRIDGE EMPTY
W2: BATTERY LOW
E2: BATTERY EMPTY
W5: PUMP TIMER
E5: END OF OPERATION,
the symbol for that Warning or Error stays in the RUN screen or
STOP screen as a Reminder.
Battery low symbol
Cartridge low symbol
Pump timer symbol
Key Lock symbol
For other Warnings or Errors, alarm information is cleared from
the display but stays in the alarm history and the event memory.
When a Warning or an Error occurs:
j
Press
f
to turn off the beeps and vibrations. The Warning or
Error code stays on the display.
j
As soon as you understand the reason for the Warning or Error,
confirm by pressing
f
again.
j
Take appropriate action, as necessary.
j
If more than one Warning and /or Error occurs simultaneously,
press
f
twice for each Warning or Error to confirm them and
turn them off.
j
After a Warning or Error, make sure that you switch your pump
back to RUN mode where appropriate.
Note
When you use your pump together with the Meter you can
also snooze and confirm an incoming Warning or Error
with the Meter. For further information see the Meter user
information.
man_05957613001_E_en.indb 152
02.03.2015 10:44:48
Содержание ACCU-CHEK Spirit Combo
Страница 8: ...man_05957613001_E_en indb 6 02 03 2015 10 43 53...
Страница 28: ...24 man_05957613001_E_en indb 24 02 03 2015 10 43 55...
Страница 64: ...60 man_05957613001_E_en indb 60 02 03 2015 10 44 15...
Страница 136: ...132 man_05957613001_E_en indb 132 02 03 2015 10 44 45...
Страница 146: ...142 man_05957613001_E_en indb 142 02 03 2015 10 44 47...
Страница 226: ...222 My notes man_05957613001_E_en indb 222 02 03 2015 10 45 07...