164
Error E6: MECHANICAL ERROR
A mechanical Error has been discovered during a routine
automatic system check and insulin delivery has stopped.
j
Press
f
twice to snooze and to confirm the Error.
j
Remove or disconnect the infusion set from your infusion site.
j
Remove the cartridge, adapter, infusion set, and battery from
your pump. Insert a new battery.
j
It is essential that you rewind the piston rod. Install a new
cartridge and infusion set. (See section
2.7
Inserting the
cartridge
). Prime the infusion set (See section
2.8
Priming the
infusion set
).
j
Reconnect the infusion set at your infusion site.
j
Put your pump in RUN mode.
If the Error E6: MECHANICAL ERROR occurs again, use the
alternative therapy plan provided by your doctor or healthcare
team. Contact your local pump support (details on inside back
cover) for more assistance.
Error E7: ELECTRONIC ERROR
An electronic Error has been discovered during a routine
automatic system check and insulin delivery has stopped.
Note
You cannot turn off and confirm an Error E7: ELECTRONIC
ERROR by pressing
f
. You must remove the battery
instead.
man_05957613001_E_en.indb 164
02.03.2015 10:44:49
Содержание ACCU-CHEK Spirit Combo
Страница 8: ...man_05957613001_E_en indb 6 02 03 2015 10 43 53...
Страница 28: ...24 man_05957613001_E_en indb 24 02 03 2015 10 43 55...
Страница 64: ...60 man_05957613001_E_en indb 60 02 03 2015 10 44 15...
Страница 136: ...132 man_05957613001_E_en indb 132 02 03 2015 10 44 45...
Страница 146: ...142 man_05957613001_E_en indb 142 02 03 2015 10 44 47...
Страница 226: ...222 My notes man_05957613001_E_en indb 222 02 03 2015 10 45 07...